Effect of the Forward Projected Model-Based Iterative Reconstruction Solution (FIRST) on
Improving Image Quality of Coronary Artery Stent in Cardiac CT Angiography
CAO Likun1,WANG Yining1,YI Yan1,WANG Yun1,MA Zhuangfei2,YAN Jing2,FU Haihong1,JIN Zhengyu1
Effect of the Forward Projected Model-Based Iterative Reconstruction Solution (FIRST) on
Improving Image Quality of Coronary Artery Stent in Cardiac CT Angiography
CAO Likun1,WANG Yining1,YI Yan1,WANG Yun1,MA Zhuangfei2,YAN Jing2,FU Haihong1,JIN Zhengyu1
Author information+
1. Department of Radiology, Peking Union Medical College Hospital, Beijing 100730, China丨
2. Canon Medical System (China) Co., Ltd., Beijing 100015, China
{{custom_zuoZheDiZhi}}
{{custom_authorNodes}}
{{custom_bio.content}}
{{custom_bio.content}}
{{custom_authorNodes}}
Show less
History+
Received
Published
2020-11-23
2021-11-10
Issue Date
2021-12-17
Abstract
Objective To explore the potential of forward projected model-based iterative reconstruction solution (FIRST) in improving
the image quality of coronary artery stents in CT angiography (CTA). Methods A total of 44 patients with 59 coronary stents who
underwent coronary CTA in our hospital were included retrospectively. The images were reconstructed with FIRST-Cardiac (FIRST-C),
FIRST-Cardiac Sharp (FIRST-CS), adaptive iterative dose reduction 3D (AIDR-3D) standard and filtered back projection (FBP)
respectively. All images were evaluated objectively, including signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) of stent
lumen, visible inner diameter of stent, and the stent lumen attenuation increase ratio (SAIR). Two radiologists visually evaluated the
image quality of the stents and performed double-blind subjective scoring with 5-point scale (1=poor, 5=excellent). The difference
between objective and subjective evaluation was compared between the four groups. Results Objective analysis of image quality
showed that there were significant differences in SNR, CNR, visible inner diameter of stent and SAIR among these four groups
(P<0.001). The SNR and CNR of FIRST-C group were significantly better than the other three groups (P<0.05)丨 the visible inner
diameter of stent in the FIRST-CS group was the largest (1.96±0.42) mm, and the SAIR was the lowest 0.32±0.27, which was
better than the other three groups. Among the subjective scores, FIRST-C and FIRST-CS images met the diagnostic requirements.
The score of FIRST-CS group was the highest, which was sequentially higher than that of FIRST-C, AIDR-3D group and FBP group.
Conclusion Compared with AIDR-3D and FBP, FIRST helps to improve the image quality of coronary stent. FIRST-C can effectively
improve the SNR. FIRST-CS has more advantages in stent lumen visualization and can provide high-quality images for clinic
CAO Likun, WANG Yining, YI Yan, et al. , {{custom_author.name_en}}et al.
Effect of the Forward Projected Model-Based Iterative Reconstruction Solution (FIRST) on
Improving Image Quality of Coronary Artery Stent in Cardiac CT Angiography[J]. China Medical Devices, 2021, 36(11): 71-75 https://doi.org/10.3969/j.issn.1674-1633.2021.11.016
{{custom_sec.title}}
{{custom_sec.title}}
{{custom_sec.content}}
References
[1] Benjamin EJ,Blaha MJ,Chiuve SE,et al.Heart disease and
stroke statistics-2017 update: a report from the American heart
association[J].Circulation,2017,135(10):e146-e603.