The Application of SNC Patient and 3DVH Software in Dose Verification of Intensity Modulated Radiotherapy Plans for Breast and Cervix Cancer

YANG Lu, ZHANG Guoqian, LUO Songgui, ZHANG Shuxu

China Medical Devices ›› 2021, Vol. 36 ›› Issue (8) : 58-62.

PDF(1440 KB)
PDF(1440 KB)
China Medical Devices ›› 2021, Vol. 36 ›› Issue (8) : 58-62. DOI: 10.3969/j.issn.1674-1633.2021.08.014
RESEARCH WORK

The Application of SNC Patient and 3DVH Software in Dose Verification of Intensity Modulated Radiotherapy Plans for Breast and Cervix Cancer

  • YANG Lu, ZHANG Guoqian, LUO Songgui, ZHANG Shuxu
Author information +
History +

Abstract

Objective To study the application of SNC Patient and 3DVH software in the dose verification of IMRT plans for breast and cervix cancer. Methods Eleven cases of breast cancer and seven cases of cervical cancer patients were selected. IMRT plans for these patients were designed by Pinnacle Treatment Planning System. The ArcCHECK phantom was chosen to verify the dose of these cases. The differences between the TPS calculated dose and measured reconstructed dose were compared by the SNC Patient and 3DVH software for every patient. Difference indicators include γ passing rate (3%/3 mm, TH=10%), target dose parameters D98%, D2%, Dmean and organ-at-risk dose parameters Dmean, etc. Results The minimum value of SNCγ passing rate for each patient was 97.4, and all of them were greater than the γ passing rate of 3DVH. The γ passing rates of each patient for the target area was greater than 90%, only the D98% difference value of the PCTV for cervix cancer patients’ was statistically significant, and the maximum difference value was less than 3%. Thirdly, the γ passing rates of each organ at risk were greater than 90%. Only the difference values for V20 of the ipsilateral lung, the Dmean of the contralateral lung and heart, the V50 of the rectum and bladder dose parameters were statistically significant, and the maximum dose difference was less than 10%. Conclusion With the help of SNC Patient software, the difference of the phantom dose distribution can be preliminaries analyzed between the TPS calculation and measurement. More specific information about the target and OAR can be provided by the 3DVH software, which was helpful for the safe implementation of the patient’s treatment plan.

Cite this article

Download Citations
YANG Lu, ZHANG Guoqian, LUO Songgui, et al. The Application of SNC Patient and 3DVH Software in Dose Verification of Intensity Modulated Radiotherapy Plans for Breast and Cervix Cancer[J]. China Medical Devices, 2021, 36(8): 58-62 https://doi.org/10.3969/j.issn.1674-1633.2021.08.014

References

[1] Cho B.Intensity-modulated radiation therapy: a review with a physics perspective[J].Radiat Oncol J,2018,36(1):1-10.
PDF(1440 KB)

686

Accesses

0

Citation

Detail

Sections
Recommended

/