Four Cases of Common Fault Maintenance Analysis of TBA-120FR Automatic Biochemical Analyzer
LI Yingjun, LI Yingchao
Author information+
Department of Clinical Laboratory, Shilin Tianqi Hospital, The Second Affiliated Hospital of Kunming Medical University,
Shilin Yunnan 652200, China
{{custom_zuoZheDiZhi}}
{{custom_authorNodes}}
{{custom_bio.content}}
{{custom_bio.content}}
{{custom_authorNodes}}
Show less
History+
Received
Published
2019-06-10
2020-02-10
Issue Date
2020-04-20
Abstract
Fully automatic biochemical analyzer is used for quantitative analysis and detection of specific substances in blood and
urine, but frequent use causes frequent failures. We summarized the causes, troubleshooting and maintenance results of four faults in
Toshiba TBA-120FR analyzer. The working principle of the automatic biochemical analyzer was analyzed. Several malfunctions were
solved, including unable to absorb the reference liquid, low test results of the patient samples, the cross contamination of the sample
results, and poor reproducibility of the results. Moreover, the device was guaranteed to be used normally after troubleshooting. We
also hope that the mentioned maintenance cases can provide reference for industry workers to repair faults of similar equipment.
LI Yingjun, LI Yingchao. , {{custom_author.name_en}}et al.
Four Cases of Common Fault Maintenance Analysis of TBA-120FR Automatic Biochemical Analyzer[J]. China Medical Devices, 2020, 35(2): 174-177 https://doi.org/10.3969/j.issn.1674-1633.2020.02.044