Three Cases of Troubleshooting for Baixiang Surgical Instrument Automatic Cleaner

LI Shu1, ZHANG Yi1, YU Yushan2

China Medical Devices ›› 2017, Vol. 32 ›› Issue (9) : 96-98.

PDF(1553 KB)
PDF(1553 KB)
China Medical Devices ›› 2017, Vol. 32 ›› Issue (9) : 96-98. DOI: 10.3969/j.issn.1674-1633.2017.09.024
MEDICAL ENGINEERING TECHNOLOGY

Three Cases of Troubleshooting for Baixiang Surgical Instrument Automatic Cleaner

  • LI Shu1, ZHANG Yi1, YU Yushan2
Author information +
History +

Abstract

Through analysis of troubleshooting cases for Baixiang QPQ500 surgical instruments automatic cleaner, maintenance methods of this machine were discussed. Statistic analysis of Baixiang QPQ500’s regular faults, including door switch failure, black screen and drainage system fault, as well as cause of these failures in our hospital between July 2012 to December 2016 were performed, and measures were adopted to restore these faults. Regular and preventive maintenance can decrease faults rate, reduce maintenance expenditure and guarantee medical devices’ normal using.

Key words

surgical instruments cleaning machine / drainage pump / circulating pump / door mechanism / overheating protection device / preventive maintenance

Cite this article

Download Citations
LI Shu, ZHANG Yi, YU Yushan. Three Cases of Troubleshooting for Baixiang Surgical Instrument Automatic Cleaner[J]. China Medical Devices, 2017, 32(9): 96-98 https://doi.org/10.3969/j.issn.1674-1633.2017.09.024

References

[1] 董辉辉.Belimed WD230清洗消毒机常见故障排除[J].中国医 疗设备,2010,25(11):135-136. [2] 孙荣江.白象QPQ550全自动清洗机门故障检修[J].医疗装 备,2014,(7):103. [3] 北京白象.白象QPQ系列全自动手术器械清洗机[J].中国医疗 器械信息,2014,8(10):100-101.