Measurement and Analysis of Oxygen Wastage of Elisee 250 Emergency Transport Ventilator

ZHENG Ji-feng, WU Hang, GUO Mi-jia, BAI Mei

China Medical Devices ›› 2017, Vol. 32 ›› Issue (6) : 90-92.

PDF(1650 KB)
PDF(1650 KB)
China Medical Devices ›› 2017, Vol. 32 ›› Issue (6) : 90-92. DOI: 10.3969/j.issn.1674-1633.2017.06.023
MEDICAL ENGINEERING TECHNOLOGY

Measurement and Analysis of Oxygen Wastage of Elisee 250 Emergency Transport Ventilator

  • ZHENG Ji-feng, WU Hang, GUO Mi-jia, BAI Mei
Author information +
History +

Abstract

Objective To measure the oxygen consumption of operating ventilator so as to accurately grasp the actual oxygen consumption of oxygen in different oxygen concentration. Methods The realtime oxygen consumption value of the breathing machine was measured under the inhaled oxygen concentration values of 21%, 30%, 40%, 50%, 60%, 70%, 80%, 90% and 100% by using the method of joining-up pipe stringing gas mass flow meter in operation of the air inlet tube on ventilator oxygen. Results The oxygen consumption of the Elisse 250 emergency transport ventilator was significantly less than LTV 1000 transport ventilator under the condition of oxygen concentration 30%~60%. Especially when the oxygen concentration was 21%, the oxygen consumption was 0, which indicated Elisee 250 emergency transport ventilator was preferred in transshipment in saving oxygen. Conclusion The oxygen consumption measurements of transport ventilator at different oxygen concentration can help medical personnel to have a correct understanding in the transport of patients with ventilator oxygen consumption. The performance testing and quality control of the transport ventilator is crucial for ensuring the safety of patients when transport of critically ill patients. Therefore, it is very necessary to perform accurate measurement and evaluation of the operation of ventilation to ensure the safety of the patients.

Key words

transport ventilator / oxygen concentration / oxygen wastage / gas storage capacity / pressure indication

Cite this article

Download Citations
ZHENG Ji-feng, WU Hang, GUO Mi-jia, et al. Measurement and Analysis of Oxygen Wastage of Elisee 250 Emergency Transport Ventilator[J]. China Medical Devices, 2017, 32(6): 90-92 https://doi.org/10.3969/j.issn.1674-1633.2017.06.023

References

[1] 程宝珍.便携式呼吸机在急诊危重患者院内安全运转的应用[J]. 中国急救复苏与灾害医学杂志,2014,9(4):379-380. [2] Grant S,Khan F,Keijzers G,et al.Ventilator-assisted preoxygenation: Protocol for combining non-invasive ventilation and apnoeic oxygenation using a portable ventilator[J].Emerg Med Australas,2016,28(1):67-72.
PDF(1650 KB)

1520

Accesses

0

Citation

Detail

Sections
Recommended

/