Low Power Design of Wearable Dynamic ECG Recorder

  • ZHENG Jun, LOU Li-gang, MAO Bin, FENG Jing-yi
Author information +
History +

Abstract

This paper introduces a power consumption design of wearable dynamic electrocardiograph (ECG) recorder with the core processor MSP430F5342 to solve the problem that working time of dynamic ECG recorder is not long. By comparing the power consumption of the chip, the MSP430F5 series chipset and the MSP430 MCU clock system are chosen, and the program of the software is used to conduct energy saving control. The low-power design can selectively shut off unused modules, which makes it possible to work in a low-power operating mode most of the time. 3 groups of power test results show that even a small lithium battery of 185 mAh can make its standby time last for 140 d, and the continuous working time can reach 7 d. Because the device uses ultra-small battery capacity, so the recorder has a small size, which makes it suitable to carry and can work for a long time. The low power design of the recorder is successful.

Key words

portable / low power consumption / wearable dynamic ECG recorder / processor / lithium battery

Cite this article

Download Citations
ZHENG Jun, LOU Li-gang, MAO Bin, et al. Low Power Design of Wearable Dynamic ECG Recorder[J]. China Medical Devices, 2016, 31(12): 113-115 https://doi.org/10.3969/j.issn.1674-1633.2016.12.031

References

[1] 陈伟伟,高润霖,刘力生,等.《中国心血管病报告2014》概要[J]. 中国循环杂志,2015,31(7):617-622. [2] 杨琦.基于MSP430超低功耗MCU的便携式心电监护仪及其 系统的研究[D].福州:福州大学,2003:5. [3] 梁航,王剑钢.基于超低功耗单片机MSP430F168的家用动态 心电记录器[J].中国组织工程研究与临床康复,2007,2(1):84- 85.
PDF(1687 KB)

1486

Accesses

0

Citation

Detail

Sections
Recommended

/