Field Application Effect Evaluation of Pulsing High Concentration Individual Oxygen
Generator on the Plateau
XIE Kangning1, HAN Wei1,2, JIANG Maogang1, RUAN Junyong1, DING Pan1, LUO Erping1
1. School of Military Biomedical Engineering, Air Force Medical University, Xi’an Shaanxi 710032, China; 2. Department of
Medical Engineering, No. 987 Hospital of PLA, Baoji Shaanxi 721000, China
Abstract:Objective To evaluate the anti-hypoxia effect of pulsing high concentration individual oxygen generator in plateau
areas. Methods We recruited 18 subjects who reached the plateau rapidly, and 14, 12 and 8 subjects who had accustomed to the
plateau at the altitudes of 3700, 4200 and 5370 m respectively, in this study. Under the resting state, their pulse oxygen saturation
(SpO2) and heart rate (HR) data were recorded when they were on the plain (1250 m), before and after oxygen inhalation on
the plateau. The results were compared by statistical methods of paired t test. Results After oxygen intake by the oxygen generator,
the average SpO2 significantly increased by 10.78%, the average HR significantly decreased by 11.06 bpm in the subjects who reached
the plateau rapidly (P<0.01). Among subjects who had accustomed to the plateau at the three different altitudes, the average SpO2
significantly increased by 5.22%, 6.33% and 9.85%, and their HR significantly decreased by 14.14, 6.83 and 15.11 bpm, respectively
(P<0.01). Conclusion The oxygen generator has remarkable anti-hypoxia effect, and can quickly relieve the hypoxia state, which
is suitable for both rapidly reached and accustomed people to the plateau.
谢康宁1,韩伟1,2,姜茂刚1,阮俊勇1,丁攀1,罗二平1. 脉冲式高浓度单兵制氧机高原实地应用效果评价[J]. 中国医疗设备, 2019, 34(5): 32-34.
XIE Kangning1, HAN Wei1,2, JIANG Maogang1, RUAN Junyong1, DING Pan1, LUO Erping1. Field Application Effect Evaluation of Pulsing High Concentration Individual Oxygen
Generator on the Plateau. China Medical Devices, 2019, 34(5): 32-34.