Study on the Cleaning and Maintenance Method of AB Suction Pipe of
GAMBRO AK Series Hemodialysis Machine
LIU Minghong1,ZHANG Haiyue2,WEI Ying2,HAN Yongshuang2,LU Xin1
1. Department of Nephrology, Jilin Municipal People’s Hospital, Jilin Jilin 132001, China丨
2. Jilin Fusheng Medical Equipment Co., Ltd., Changchun Jilin 130607, China
Abstract:Objective To discuss the cleaning and maintenance methods of AB suction pipe of GAMBRO AK series hemodialysis
machine. Methods The heated dialysate was used to treat the AB suction pipe, and the ion content in the crystallization and residual
liquid of AB suction pipe with different water temperature, soaking time and frequency was compared to determine the optimum
treatment conditions. Results More crystals and residues on the dialysate AB tube were eluted with the increase of water temperature
and the prolongation of time, and there was a significant difference (P<0.05). There was no significant difference when the water
temperature exceeded 80 ℃ and the time exceeded 60 s (P>0.05). There was no significant difference in the crystals appearing in
AB pipette, as well as the sodium, potassium, and calcium ions in the residual liquid when the treatment was carried out after the
first, second and third shifts of dialysis (P>0.05). Conclusion The optimal treatment condition is to soak the AB suction pipe in
80℃ dialysis water for 60 s after daily dialysis. The method has the advantages of simple operation and low cost with good treatment
effect. It can reduce the failure rate of dialysis machine.
刘铭红1,张海悦2,魏滢2,韩永双2,鲁新1. 金宝AK系列血液透析机AB吸液管清理维护方法的研究[J]. 中国医疗设备, 2021, 36(11): 128-131.
LIU Minghong1,ZHANG Haiyue2,WEI Ying2,HAN Yongshuang2,LU Xin1. Study on the Cleaning and Maintenance Method of AB Suction Pipe of
GAMBRO AK Series Hemodialysis Machine. China Medical Devices, 2021, 36(11): 128-131.