Effect of SFOV on Water Equivalent Diameter of Size-Specific Dose Estimation in Adult Chest CT
HU Qinqin, ZHANG Qin, LIU Xi, ZHANG Dechuan, YANG Huiping, KONG Lina, ZHANG Hao, LI Xiujian
Author information+
Department of Radiology, Chongqing Traditional Chinese Medicine Hospital, Chongqing 400021, China
{{custom_zuoZheDiZhi}}
{{custom_authorNodes}}
{{custom_bio.content}}
{{custom_bio.content}}
{{custom_authorNodes}}
Show less
文章历史+
收稿日期
出版日期
2021-09-01
2021-10-10
发布日期
2021-11-04
摘要
目的 探讨扫描野(Scanning Field of View,SFOV)对成人胸部CT水当量直径的体型特异性剂量估算(Water
Equivalent Diameter of Size-Specific Dose Estimates,SSDEWED)的影响。方法 回顾性分析2020年7至8月我院行胸部CT扫描
的162例成人受检者,选取原始序列横断位中间层面图像,勾画圆形感兴趣区(Region of Interest,ROI),测量ROI的CTROI
值、面积AROI,记录受检者的性别、年龄、容积剂量指数(Volume CT Dose Index,CTDIvol)、剂量长度乘积(Dose Length
Product,DLP)和SFOV的数值,计算每例受检者中间层面的水等效直径(Water Equivalent Diameter,WED)、转换因子
(fsize·WED)和SSDEWED。利用多元回归分析性别、年龄、CTROI、AROI、CTDIvol、DLP和SFOV与SSDEWED的相关性。结果 性
别、年龄、AROI、CTDIvol、DLP和SFOV与SSDEWED有显著差异(P<0.05),CTROI与SSDEWED无显著差异(P>0.05),多元
回归分析SSDEWED与SFOV呈0.06倍正线性相关。结论 SFOV对成人胸部CT的SSDEWED影响较大,且呈0.06倍正线性相关,
在临床扫描中精准调节SFOV可以降低成人受检者胸部CT辐射剂量。
Abstract
Objective To investigate the effect of scanning field of view (SFOV) on water equivalent diameter of size-specific dose
estimates (SSDEWED) in adult chest CT. Methods A total of 162 adult patients who underwent chest CT scan in our hospital from
July to August in 2020 were retrospectively analyzed. The middle level images of the original sequence were selected, and the
circular region of interest (ROI) was drawn. The CT value and area of ROI were measured. The gender, age, volume CT dose index
(CTDIvol), dose length product (DLP) and SFOV were recorded. The water equivalent diameter (WED), fsize·WED and SSDEWED in the
middle layer of each patient were calculated. Multiple regression analysis was used to analyze the correlation between gender, age,
CTROI, AROI, CTDIvol, DLP, SFOV and SSDEWED. Results Gender, age, AROI, CTDIvol, DLP and SFOV were significantly different
from SSDEWED (P<0.05),but CTROI were no significantly different from SSDEWED (P<0.05). Conclusion SFOV has a significant
influence on the SSDEWED of chest CT scan, with a positive linear correlation of 0.06 times. Precisely adjusting SFOV can reduce the
chest CT radiation dose of adult patients in clinical scan.
HU Qinqin, ZHANG Qin, LIU Xi, et al. , {{custom_author.name_en}}et al.
Effect of SFOV on Water Equivalent Diameter of Size-Specific Dose Estimation in Adult Chest CT[J].
China Medical Devices, 2021, 36(10): 140-143 https://doi.org/10.3969/j.issn.1674-1633.2021.10.032