1. Department of Physical Testing, Henan Medical Equipment Inspection Institute, Zhengzhou Henan 450000, China丨 2. Department
of Physical Testing, Henan Medical Equipment Inspection and Testing Engineering Technology Research Center, Zhengzhou
Henan 450000, China丨 3. Department of Research and Development, Henan Camel Medical Equipment Research Institute Co., Ltd,
Xinxiang Henan 453400, China
Abstract:Objective To improve the sterilization efficiency of medical protective masks, the effect of electron beam irradiation
sterilization on the filtration efficiency of medical protective masks was studied. Methods According to GB 19083-2010 technical
requirements for medical protective masks, testing filtration efficiency of the samples produced by ten enterprises (A~J) and
sterilized by different doses (3 kGy, 4 kGy, 5 kGy, unpasteurized) of electron beam irradiation with TSI 8130 automated filter
tester. Results The filtration efficiency of the samples was unqualified, and the electron beam irradiation sterilization with the dose of
3 kGy, 4 kGy and 5 kGy was not suitable for the medical protective mask. Conclusion It is not feasible to carry out rapid sterilization
of medical protective masks by electron beam irradiation.
张峻梓1,2,邓杰1,2,朱政辉1,2,王国锋3. 电子束辐照灭菌对医用防护口罩过滤效率的影响[J]. 中国医疗设备, 2021, 36(7): 134-136.