Cases of Failure Maintenance of GE730 Color Doppler Ultrasound Diagnostic Instrument
CHEN Jianjun1,DONG Rui2a,HAN Yefeng2b
1. Department of Equipment, Cixi Third People’s Hospital, Cixi Zhejiang 315300, China丨
2. Radiation Therapy Center, Cixi People’s Hospital, Cixi Zhejiang 315300, China
Abstract:Color doppler ultrasound equipment is a kind of medical instrument with safety, fast imaging and high diagnostic value,
which has a high social and economic benefits in clinical use. Due to its complex structure and high precision, various failures will
inevitably occur in the process of use. In this paper, the basic structure of GE 730 color doppler ultrasound was analyzed. And the
fault judgment basis and the specific repair method ware expounded in detail through two fault cases. The circuit schematic diagram
was drawn by the actual circuit board, and then the circuit analysis and detection were carried out to determine the fault components,
and finally the maintenance was completed by replacing the electronic components. This maintenance method can achieve the lowest
maintenance cost, save a lot of maintenance expenses for hospitals, and enhance the professional value of clinical engineers.
陈建均1,董蕊2a,韩叶锋2b. GE 730彩色多普勒超声诊断仪故障维修案例[J]. 中国医疗设备, 2021, 36(5): 184-186.