Abstract:According to the statistics of the fault data of our hospital, the clinical failures of D3 defibrillator were mainly concentrated in the front shell module, multi-parameter module and power module. The main clinical manifestations of failures were failure to boot, failure to charge and discharge, no display on black screen, and no ECG waveform, which these four types accounted for about 64% of total failures. Rapid troubleshooting and maintenance can not only ensure the intact rate of defibrillator, but also reduce maintenance costs, while effectively ensuring the safety of equipment.
魏昊业,宗会迁,柳青,杨存. 迈瑞D3除颤监护仪常见故障分析与维修[J]. 中国医疗设备, 2020, 35(9): 173-175.
WEI Haoye, ZONG Huiqian, LIU Qing, YANG Cun. Failure Analysis and Maintenance of Mindray D3 Defibrillator Monitor. China Medical Devices, 2020, 35(9): 173-175.