Troubleshooting and the Maintenance for PARAMOUNT
CA-38G Multi-functional Monitoring Bed
NIU Weiminga,XIANG Siweib,ZHENG Kunb
a. Department of Radiology| b. Department of Medical Equipment, The Children’s Hospital, Zhejiang University School
of Medicine, National Clinical Research Center for Child Health, Zhejiang Hangzhou 310052, China
Abstract:PARAMOUNT CA-38G multi-functional monitoring bed is a kind of ICU monitoring bed widely used in our hospital. The equipment is safe and reliable, and can effectively meet the using requirement of clinical. Two typical fault maintenance cases were introduced by this paper. The first one is motor failure due to relay failure. And the second one is the equipment can’t be used normally which caused by fault power control board. Through observing the fault phenomenon and analyzing the reason step by step. Finally, by replacing the new motor relay, zener diode, the failures were finally repaired successfully. It also leads to the importance of the usual maintenance of the monitoring bed and provides a guarantee for the normal operation of the hospital through rapid troubleshooting.