Abstract:To standardize the detection of the maximum allowable error of infrared thermometer for manufacturing enterprises and testing institutions, this paper proposes strategies based on the requirements of GB/T 21417.1-2008, GB/T 19146-2010 and GB/T 19665-2005. This article introduced the test techniques and methods of the maximum allowable error, and summarized the test precautions, and also found that the measurement distance is different, the measuring optical axis does not coincide with the black body cavity axis, and the maximum allowable error changes significantly. The correct test method of maximum allowable error proposed in this paper can effectively avoid detection error and improve detection efficiency. Moreover, this method is helpful to ensure the safety and effectiveness of products before marketing, and to standardize and guide the detection of medical devices.
邓振进,吴碧涛,刘鹏举,彭再明,彭明霞. 红外测温仪最大允许误差测试方法的研究[J]. 中国医疗设备, 2020, 35(9): 84-87.
DENG Zhenjin, WU Bitao, LIU Pengju, PENG Zaiming, PENG Mingxia. Test Method of Maximum Allowable Error of Infrared Thermometer. China Medical Devices, 2020, 35(9): 84-87.