Working Principle and the Troubleshooting of SHINVA Super6000
Cleaning and Disinfecting Apparatus
QIU Kefeng, ZHENG Kun, ZHENG Caixian
Department of Medical Equipment, The Children’s Hospital of Zhejiang University School of Medicine, National Clinical Research
Center for Child Health, Hangzhou Zhejiang 310003, China
Abstract:Firstly, this paper analyzed the structure and working principle of Xinhua Super6000 cleaning sterilizer. Then, taking three
kinds of faults of the equipment, such as “front door motor protection” “water not ready” “air leakage in the process of equipment
static and operation”, and the maintenance process as examples, the test method and troubleshooting process of relevant faults were
introduced in detail. The equipment was restored to normal operation by replacing the PLC connecting line, pneumatic valve and
front door cylinder. It is expected that this paper can provide a reference for the rapid repair of faults and regular inspection system,
so that the equipment operation is more stable and better for clinical services.
裘科烽,郑焜,郑彩仙. 新华Super6000型清洗消毒器工作原理和故障维修[J]. 中国医疗设备, 2020, 35(8): 171-173.
QIU Kefeng, ZHENG Kun, ZHENG Caixian. Working Principle and the Troubleshooting of SHINVA Super6000
Cleaning and Disinfecting Apparatus. China Medical Devices, 2020, 35(8): 171-173.