Abstract:This study aimed to ensure the safety of medical device products and to improve the effectiveness and feasibility of
conducted susceptibility evaluation of medical electrical equipment. Based on the standard GB/T17626.6-2017, YY0505-2012, and
YY0607-2007, this paper analyzed the effect of current clamp injection at different ports on the interference of medical electrical
equipment. We used the nerve and muscle stimulator equipment as the test equipment. The results showed that the experimental
injection mode of single cable (one port) current clamp injection was more significant. It is expected to form a reference and
specification for the electromagnetic compatibility test of nerve, muscle stimulator medical electrical equipment and improve the
reproducibility of the test results. Meanwhile, it may guide the manufacturers of nerve and muscle stimulators of medical electrical
equipment in the production phase of anti-interference design.
王美,公翠娟,张淑溢,杨姝,李庆雨,徐志洲. 医用电器设备的射频场感应的传导抗扰度测试方法分析[J]. 中国医疗设备, 2020, 35(3): 25-27.
WANG Mei, GONG Cuijuan, ZHANG Shuyi, YANG Shu, LI Qingyu, XU Zhizhou. Analysis of Induction of the Radio-Frequency Fields Conduction Immunity
Test Method of Medical Electrical Equipment. China Medical Devices, 2020, 35(3): 25-27.