Abstract:To provide reference experience for common faults in repairing ordinary mobile DR, three high-frequency faults of
MUX-100DJ mobile DR were analyzed and overhauled. Based on the operation principle of the equipment, we analyzed the failure
of the ball cross arm travel speed, reported the fault code “F32” or “F33” in the startup or preparation state and reported the “d1”
error code one by one, and quickly eliminated the fault. We also recognize that the industry workers not only need to quickly solve
the equipment failure caused by aging, wear and tear of components, and lack of maintenance factors, but also combine periodic and
preventive maintenance work.
喻志才,蒋元林. MUX-100DJ移动DR故障分析及处理三例[J]. 中国医疗设备, 2020, 35(2): 181-184.
YU Zhicai, JIANG Yuanlin. Three Cases of Troubleshooting for MUX-100DJ Mobile DR Machine. China Medical Devices, 2020, 35(2): 181-184.