Abstract:Objective By describing three failures and maintenance processes of Elekta Synergy linear accelerator, the ideas and
methods of related problems were introduced, which could provide references for colleagues. Methods Some errors occurred,
such as Low dose rate, Water temperature alarm and Touch guard interlock, based on the equipment failure phenomenon,
software error messages, and combined with circuit diagram analysis, related faulty components were replaced. Results The
system structure was well acquainted with, relevant experience was summarized, and the loss caused by equipment shutdown was
avoided. Conclusion The linear accelerator system is complex and complicated with several failures. As long as the correct
maintenance ideas are mastered and the failure reasons are analyzed, the maintenance technology can be improved, and
the operating costs can be saved.
刘立洁a,张鹏a,李书红b,徐英c,耿西亮a. Synergy直线加速器故障维修三例[J]. 中国医疗设备, 2019, 34(6): 170-172.