心室辅助装置溶血检测及关键问题研究

侯晓旭,李澍,王浩,罗维娜,王权,任海萍

中国医疗设备 ›› 2018, Vol. 33 ›› Issue (9) : 15-18.

PDF(1593 KB)
PDF(1593 KB)
中国医疗设备 ›› 2018, Vol. 33 ›› Issue (9) : 15-18. DOI: 10.3969/j.issn.1674-1633.2018.09.004
专论

心室辅助装置溶血检测及关键问题研究

  • 侯晓旭,李澍,王浩,罗维娜,王权,任海萍
作者信息 +

Study on Testing and Key Problems of Hemolysis Evaluation in Ventricular Assist Device

  • HOU Xiaoxu, LI Shu, WANG Hao, LUO Weina, WANG Quan, REN Haiping
Author information +
文章历史 +

摘要

目的 溶血是评价心室辅助装置安全性的重要方面,以溶血指数作为评价对象。其物理意义是反映输出100 L压积标 准化血液产生的血浆游离血红蛋白含量,是目前世界上公认的体外评价血泵溶血性能的标准。方法 本文参照ASTM F1841- 97标准规定的原则,对溶血试验进行了详细的设计,搭建了检测平台,进行了溶血指数测量,并针对标准中未进行详细说 明的血液初始质量要求及游离血红蛋白检验方法进行研究和实现。结果 成功设计并搭建了溶血试验检测平台,基于此检 测平台成功对一台血泵的溶血指数进行了测试,表明了检测平台的有效性。结论 采用ASTM F1841-97标准,搭建了检测平 台,进行了溶血指数测量,明晰了实现方法和细节,希望供研发和评价人员借鉴。

Abstract

Objective The physical significance of the hemolytic index is to reflect the content of free hemoglobin in 100 L output hematocrit and to reflect the safety of the ventricular assist device content, which is recognized as the standard in the world to evaluate the hemolysis performance of the blood pump. Methods Based on the introduction of ASTM F1841-97, this paper focused on the refinement and discussion of the parts that not specified in the standard. It included the initial blood quality requirement and the free hemoglobin test method. Results The detecting platform for hemolysis testing was successfully designed and built, and was used for the test of hemolysis performance of one blood pump, indicating the effectiveness of the platform. Conclusion Based on the introduction of ASTM F1841-97, we have established a detecting platform to realize the hemolysis testing for blood pump and explicate the test methods and details, which would provide some reference for the researcher and tester.

关键词

溶血指数 / 血泵 / 剪切力 / 心室辅助装置

Key words

haemolytic index / blood pump / shear force / ventricular assist device

引用本文

导出引用
侯晓旭, 李澍, 王浩, . 心室辅助装置溶血检测及关键问题研究[J]. 中国医疗设备, 2018, 33(9): 15-18 https://doi.org/10.3969/j.issn.1674-1633.2018.09.004
HOU Xiaoxu, LI Shu, WANG Hao, et al. Study on Testing and Key Problems of Hemolysis Evaluation in Ventricular Assist Device[J]. China Medical Devices, 2018, 33(9): 15-18 https://doi.org/10.3969/j.issn.1674-1633.2018.09.004
中图分类号: R318.11   

参考文献

[1] Foo D,Walker BD,Kuchar DL,et al.Left ventricular mechanical assist devices and cardiac device interactions: an observational case series[J].PACE,2010,32(7):879-887. [2] Meyer AL,Slaughter MS.The total artificial heart[J].Panminerva Med,2011,53(3):141-154.

基金

国家重点研发计划(2016YFC0103201)。