Analysis of CT Performance Test Results of Medical Institutions in a Region
YAO Guoqing1a, DING Qin1b, MI Yongwei1a, ZHENG Xiaoxi1a, LI Tao1a, TAN Xianjian1a, DING Zhaofei2
1.a.Department of Medical Engineering; b.Department of Ophtalmology, Wuhan General Hospital of Chinese People’s Liberation
Army, Wuhan Hubei 430070, China; 2.School of Graduate, Southern Medical University, Guangzhou Guangdong 510515, China
Abstract:Objective To understand the quality of clinical application of CT through the analysis of CT performance test results of
medical institutions. Methods The main performance indicators of 39 CT selected in 9 medical institutions in a region were tested
for quality control according to national and industry standards. Results Most CT performance status were excellent. The noise pass
rate was 97.4%, the measurement index pass rate was 92.3%, and the remaining test indicators pass rate reached 100%. Conclusion
CT performance testing is the premise and technical guarantee for its quality of clinical application. In order to better ensure the level
of performance of medical units at all levels of CT, we need to strengthen the equipment management consciousness, improve the
equipment quality control standard specification and share of quality control information, etc.
姚国庆1a,丁琴1b,米永巍1a,郑小溪1a,李涛1a,谭先健1a,丁朝飞2. 某地区医疗机构CT性能状态检测结果分析[J]. 中国医疗设备, 2018, 33(5): 97-99.