Abstract:CT has highly strict requirements both in software and hardware in use, many software problems affect the use of it
directly. This paper mainly introduced problems of CT software in use, including unable burning of patient information passed to
post processing workstation, imagine loss, slow reconstruction and unable to scan, these problems could be solved by database
maintenance and reconstruction. CT operator and manger should pay attention to protect database, process patient data regularly,
archive data and ensure its’ normal use.
杨涛a,向治平a. 西门子Somatom Defintion Flash CT数据库故障分析3例[J]. 中国医疗设备, 2018, 33(2): 102-104.
YANG Taoa, XIANG Zhipingb. Three Cases of Troubleshooting for Siemens Somatom Defintion Flash CT Database. China Medical Devices, 2018, 33(2): 102-104.