HU Hong-bo1, HU Ying1, QIU Zhi-fen2, LIU Hong-xiang1
1.Radiation Room, Institute for Drug and Instrument Control, Joint Logistics Department of Guangzhou Military Area Command, Guangdong Guangzhou 510080, China; 2.Department of Nuclear Medicine, Meizhou People’s Hospital, Meizhou Guangdong 514031, China
Abstract:ObjectiveTo design a respiratory motion artifact correction performance test phantom which is used to quantitatively analyze the operating results of PET/CT respiratory gating system. Methods This thoracic cavity model was designed based on the image quality testing body model of American National Electrical Manufacturers Association, on which the inner diameter of the same aperture was used to insert content. The contents were internal cylinders that contain different diameters of hollow ball. Fluorodeoxyglucose solution was perfused into the hollow balls to simulate the size of lung cancer. Through a high-precision motor to control the elevation of the chest phantom, and simulate chest expansion and contraction of human respiratory; another high-precision motor was used to control the rotation, in and out of tumor mimics and simulated trajectories lung cancer with chest breathing.ResultsCompared with the standard image, preliminary experimental results indicated that the tumor deviation on three dimensions were 12.42 mm (axial), 13.36 mm (coronary) and 12.28 mm (sagittal) respectively, which was consistent with the clinical research. The results showed that the respiratory motion artifact correction performance test phantom could quantitatively evaluate the breathing artifact correction effects of respiratory movement control methods.ConclusionThis movable phantom is simple, and is not subject to the operational experience of clinical staffs and individual differences of experimental subject. The use of high-precision motor to control mold body movement reduces the impact of system error during actual operation and random error, while ensuring the reproducibility of the experiment.
胡红波1,胡莹1,丘志芬2,刘鸿翔1. 一种用于PET/CT呼吸运动伪影校正性能测试的动态体模设计[J]. 中国医疗设备, 2017, 32(1): 30-33.
HU Hong-bo1, HU Ying1, QIU Zhi-fen2, LIU Hong-xiang1. Movable Phantom Design for Performance Test of PET/CT Respiratory Motion Artifact Correction. China Medical Devices, 2017, 32(1): 30-33.
Chen WQ,Zhang SW,Li LD,et al.An analysis of lung cancer incidence and mortality from 30 cancer registries in China, 1998-2002[J].Bull Chin Cancer,2006,15(9):570-574.
Chen WQ,Zhang SW,Li LD,et al.An analysis of lung cancer incidence and mortality from 30 cancer registries in China, 1998-2002[J].Bull Chin Cancer,2006,15(9):570-574.
[2]
Chen WQ,Zhang SW,Li LD,et al.An analysis of lung cancer incidence and mortality from 30 cancer registries in China, 1998-2002[J].Bull Chin Cancer,2006,15(9):570-574.