Abstract:The sterilization parameters offered by device manufacturer can meet the sterilization effect
of monitoring indicators, but the wet package can’t be avoided. To explore the optimum program for
rigid container in pulsing pre-vacuum pressure steam sterilization, different specifications of the rigid
container were loaded as required, on the basis of sterilization parameters offered by the manufacture,
according to the technical specification of the sterilization and supply center and relevant literature
reports, the sterilization effect monitoring and the presence of wet package were checked to make
reasonable adjustments of sterilizer application settings. After two adjustments, wet package rate was
lowered to 1%, and the qualification rate of sterilization articles reached 98.8%.
董美丽,吕润丽. 硬质容器在脉动预真空压力蒸汽灭菌中的最佳程序选择[J]. 中国医疗设备, 2016, 31(12): 110-112.
DONG Mei-li, LV Run-li. Selection of Optimum Program for Rigid Container in Pulsing Pre-vacuum
Pressure Steam Sterilization. China Medical Devices, 2016, 31(12): 110-112.