数字乳腺X线摄影系统成像参数优化研究
闵高瑜1a ,胡益斌1a ,金锡军1a ,李大鹏1b ,李明明2
南京医科大学第一附属医院
Preliminarily Investigation of Optimization of Imaging Parameters in Digital Mammography System
MIN Gao-yu1a , HU Yi-bin1a , JIN Xi-jun1a , LI Da-peng1b , LI Ming-ming2
1. a. Department of Clinical Engineering;
b. Department of Radiology, The First
Affiliated Hospital of Nanjing Medical
University, Nanjing Jiangsu 210029,
China; 2. Jiangsu Institute of Metrology,
Nanjing Jiangsu 210029, China
摘要 目的 本文对数字乳腺X线摄影系统中成像参数的最优化进行初步研究。方法 采用4 cm厚的PMMA乳腺模体,选择不同的靶/滤过组合、管电压和管电流时间乘积,进行手动成像。利用ImageJ软件中的qa-distri插件得到图像信号区域和背景区域的像素均值和标准差,以此计算图像的对比度噪声比。然后根据对比度噪声比和平均腺体剂量两个参数,评估靶/滤过组合、管电压和管电流时间乘积对成像的影响。结果 对于厚度为 4 cm的乳房,使用钼/铑组合,在得到同等质量图像的情况下,受检者吸收的平均腺体剂量比使用钼/钼组合低。结论,在数字乳腺X线摄影系统临床使用过程中,要选择适宜的曝光参数,既能保证图像质量,又能降低受检者的辐射风险。
关键词 :
数字乳腺X线摄影系统 ,
钼/铑组合 ,
钼/钼组合 ,
对比度噪声比 ,
平均腺体剂量
Abstract :Objective Optimizing imaging parameters in digital mammography system was preliminarily
investigated in this paper. Methods The 4-cm-thick PMMA (Poly-Methyl-Meth-Acrylate) breast
simulating phantom was used, and two different combinations of anode/filter material and a wide range
of peak tube voltage and tube current-time product were employed for manually imaging. The images
were analyzed by using a set of plug-in applications in the ImageJ software for acquisition of the pixel
mean value and standard deviation measurements so as to calculate contrast noise ratio. Then, the impact
of combination of anode/filter material, peak tube voltage and tube current-time product on images was
assessed by using contrast noise ration and mean glandular dose.
Key words :
digital mammography system
combination of Mo/Rh
combination of Mo/Mo
contrast to noise ratio
mean glandular dose
收稿日期: 2015-10-15
作者简介 :
[1]
Ferlay J,Shin HR,Bray F,et al.Cancer incidence and mortality
worldwide:IARC Cancer Base No.10.GLOBOCAN 2008[M].
Lyon:International Agency for Research on Cancer,2010.
and digital mammography:a comparison of breast cancer visibility
subtle mammographic findings[J].Eur Radiol,2008,18(12):2817-
2825.
[3]
Ragozzino M.Optimal photon energies with respect to absorbed
dose for visualization of soft tissue masses with adipose
dose for visualization of soft tissue masses with adipose
tissue[J].Med Phys,1982,9(4):493-496.
[4]
Bernhardt P,Batz L,Ruhrnschopf E,et al.Spatial frequencydependent
signal-to-noise ratio as a generalized measure of
image quality[J].Proc of the SPIE,2005,5745:407-418.
[5]
Bottigli U,Bolosio B,Masala G,et al.Effect of different
[5]
Bottigli U,Bolosio B,Masala G,et al.Effect of different
spectral distributions to image a contrast detail phantom in the
mammography energy range[J].Nuovo Cimento della Societa
Italiana di Fisica,2006,29(2):215-228.
[6]
Oliva P,Golosio B,Stumbo S,et al.Compact X-ray sources for
quality[J].Med Phys,2009,36(11):5149-5161.
quality[J].Med Phys,2009,36(11):5149-5161.
Thomson source for mammography[J].Nuclear Instrumments
Thomson source for mammography[J].Nuclear Instrumments
and Methods in Physics,2010,615(1):93-99.
image quality[J].Proc of the SPIE,2005,5745:407-418.
image quality[J].Proc of the SPIE,2005,5745:407-418.
[5]
Bottigli U,Bolosio B,Masala G,et al.Effect of different
spectral distributions to image a contrast detail phantom in the
spectral distributions to image a contrast detail phantom in the
mammography energy range[J].Nuovo Cimento della Societa
Italiana di Fisica,2006,29(2):215-228.
Thomson source for mammography[J].Nuclear Instrumments
[2]
Andersson I,Ikeda DM,Zackrisson S,et al.Breast tomosynthesis
and digital mammography:a comparison of breast cancer visibility
subtle mammographic findings[J].Eur Radiol,2008,18(12):2817-
mammographic applications:Monte Carlo simulations of image
[1]
Ferlay J,Shin HR,Bray F,et al.Cancer incidence and mortality
[1]
Ferlay J,Shin HR,Bray F,et al.Cancer incidence and mortality
worldwide:IARC Cancer Base No.10.GLOBOCAN 2008[M].
subtle mammographic findings[J].Eur Radiol,2008,18(12):2817-
2825.
2825.
[3]
Ragozzino M.Optimal photon energies with respect to absorbed
[3]
Ragozzino M.Optimal photon energies with respect to absorbed
dose for visualization of soft tissue masses with adipose
tissue[J].Med Phys,1982,9(4):493-496.
tissue[J].Med Phys,1982,9(4):493-496.
[4]
Bernhardt P,Batz L,Ruhrnschopf E,et al.Spatial frequencydependent
[4]
Bernhardt P,Batz L,Ruhrnschopf E,et al.Spatial frequencydependent
signal-to-noise ratio as a generalized measure of
signal-to-noise ratio as a generalized measure of
Lyon:International Agency for Research on Cancer,2010.
[2]
Andersson I,Ikeda DM,Zackrisson S,et al.Breast tomosynthesis
and BIRADS classification in a population of cancers with
[6]
Oliva P,Golosio B,Stumbo S,et al.Compact X-ray sources for
mammographic applications:Monte Carlo simulations of image
[7]
Oliva P,Bacci A,Bottigli U,et al.Start-to-end simulation of a
and Methods in Physics,2010,615(1):93-99.
and Methods in Physics,2010,615(1):93-99.
worldwide:IARC Cancer Base No.10.GLOBOCAN 2008[M].
Lyon:International Agency for Research on Cancer,2010.
[2]
Andersson I,Ikeda DM,Zackrisson S,et al.Breast tomosynthesis
and digital mammography:a comparison of breast cancer visibility
and BIRADS classification in a population of cancers with
and BIRADS classification in a population of cancers with
mammography energy range[J].Nuovo Cimento della Societa
Italiana di Fisica,2006,29(2):215-228.
[6]
Oliva P,Golosio B,Stumbo S,et al.Compact X-ray sources for
mammographic applications:Monte Carlo simulations of image
quality[J].Med Phys,2009,36(11):5149-5161.
[7]
Oliva P,Bacci A,Bottigli U,et al.Start-to-end simulation of a
[7]
Oliva P,Bacci A,Bottigli U,et al.Start-to-end simulation of a
[1]
蔡威,郭良栋,蔡南阳,祁子欣. 瓦里安Clinac IX直线加速器故障维修2例 [J]. 中国医疗设备, 2019, 34(3): 177-178.
[2]
张灿. 医科达Synergy医用电子直线加速器CON A故障分析与维修 [J]. 中国医疗设备, 2019, 34(3): 170-173.
[3]
郑旭海1 ,谭庭强2 ,李黎2 . Synergy直线加速器水冷系统故障维修二例 [J]. 中国医疗设备, 2019, 34(3): 168-169.
[4]
杨忠东,陈自谦,钟群,肖慧,许尚文,付丽媛. TRACERlab Fx-FDG全自动化学合成器的故障处理及日常质量控制 [J]. 中国医疗设备, 2019, 34(2): 13-16.
[5]
陈蕾蕾,陈自谦,许尚文,肖慧,钟群,付丽媛,钱根年. PET/CT显像常见伪影及质量控制 [J]. 中国医疗设备, 2019, 34(2): 17-20.
[6]
陈杰,付丽媛,钟群,肖慧,许尚文,赵春雷,钱根年,陈自谦. 双时相PET/CT检查技术在膀胱癌诊断中的价值 [J]. 中国医疗设备, 2019, 34(2): 21-24.
[7]
陈振,陈自谦,钟群,肖慧,许尚文,付丽媛,陶超超. 基于体重优化CT扫描条件对PET/CT成像的CT辐射剂量与图像质量的影响研究 [J]. 中国医疗设备, 2019, 34(2): 25-27.
[8]
钱根年,陈自谦,钟群,肖慧,许尚文,付丽媛. PET/CT成像系统的质量控制 [J]. 中国医疗设备, 2019, 34(2): 1-4.
[9]
徐磊1a ,孟庆乐1a ,杨瑞1a ,钱鑫宇1a ,蒋红兵1b, 2 . uMI 780 PET/CT NEMA性能指标测试研究 [J]. 中国医疗设备, 2019, 34(1): 14-17.
[10]
雷宏昌,马蕾杰,刘冰,陈鹏,刘晗. IC Profiler性能研究 [J]. 中国医疗设备, 2019, 34(1): 49-51.
[11]
魏绪国a ,李修磊a ,王宏英a ,张明臣b ,王绪刚a . 医科达Precise型直线加速器运行情况的分析与研究 [J]. 中国医疗设备, 2019, 34(1): 117-120.
[12]
邵勤1 ,方舸1 ,严郁1 ,朱伟1 ,成志标2 ,李天鹏1 . 基于动态压力检测模块的高精度医用微量注射泵的研制 [J]. 中国医疗设备, 2018, 33(11): 40-42.
[13]
邓永锦a ,欧阳斌a ,王振宇a ,肖振华a ,黄伯天a ,黄镜先a ,李宏行b ,包勇a . Varian Novalis Tx医用直线加速器HD120MLC漏透射特性的测定及其临床影响研究 [J]. 中国医疗设备, 2018, 33(10): 49-53.
[14]
于广浩1 ,高杨1 ,徐建忠1 ,李莲娣2 . 程控X线机故障维修及其在教学中的应用 [J]. 中国医疗设备, 2018, 33(10): 107-109.
[15]
黄靖,潘克新,徐太祥,尹军. 德国SIRONA牙科曲面断层X线机检修 [J]. 中国医疗设备, 2018, 33(9): 100-102.