Abstract:Madison X6 color doppler ultrasound is widely used in small and medium-sized hospitals, and various faults occur
frequently. This paper introduced the structure principle of Madison X6, expounded three classic cases , including black screen when
starting up, no response when starting up, split-screen crash when self-checking, and self-maintenance of multiple different faults
caused by the same part. In the absence of technical manual conditions, by independently drawing the equipment principle structure
diagram, the equipment self-test process and operation status of the equipment software and hardware during self-test were analyzed.
By checking the integrated components and unit circuit information of the functional circuit board of the equipment, the equipment
were repaired by replacing components and cleaning. On the basis of maintenance analysis and summary, this paper provides
reference for engineers to eliminate and maintain faults, save maintenance and use costs, and improve the equipment integrity rate.
黄升云,万洲. 麦迪逊X6彩超结构原理与维修案例分析[J]. 中国医疗设备, 2022, 37(2): 179-182.
HUANG Shengyun, WAN Zhou. Case Analysis on Structure Principle and Maintenance of Madison X6 Color Doppler Ultrasound. China Medical Devices, 2022, 37(2): 179-182.