a. Department of Radiology| b. Department of Medical Engineering and Information, Qilu Hospital (Qingdao),
Cheeloo College of Medicine, Shandong University, Qingdao Shandong 266035, China
Abstract:Objective To explore the causes of unqualified multi-parameter monitor and provide guidance for multi-parameter monitor preventive maintenance. Methods According to the requirements of national metrology and the product specifications of multi-parameter monitor, FLUKE Prosim8 vital signs simulator was used to detect the common functions of 43 multi-parameter monitors in emergency department and anesthesiology department. Results The qualified rate of 43 multi-parameter monitors was 86.05%, the qualified rate of model A was 79.3%, and the qualified rate of model B and C was 100%, the main reason for the unqualified of the multi-parameter monitor was internal air leakage in the non-invasive blood pressure measurement. Conclusion According to the test quality control data, the discussion of this paper plays a role in the introduction of ideas and methods for the peer reference for making the regular preventive maintenance plan, reducing the equipment failure rate, and ensuring the equipment in a good state of operation.
任国荣a,祝树森b,孟祥水a,庄严b,姜文娟b,杨康b. 多参数监护仪质量控制与数据分析[J]. 中国医疗设备, 2020, 35(9): 74-74.
REN Guoronga,ZHU Shusenb,MENG Xiangshuia,ZHUANG Yanb,JIANG Wenjuanb,YANG Kangb. Quality Control and Data Analysis of Multi-Parameter Monitor. China Medical Devices, 2020, 35(9): 74-74.