Abstract:Electronic endoscope is becoming popular in clinical application. As a precise medical device, there were errors happened frequently and lead to high maintenance cost. Through analyzing structure and working principle of electronic endoscope, we summarized endoscopic operation of common faults and the cause of the problem, and proposed a series of measures before, during and after the use of endoscope. This work may help reduce the endoscopic failure rate, prolong service life, reduce maintenance cost, and provide reference for endoscopic operation and maintenance personnel.
李文莲,朱慧. 电子内窥镜常见故障分析及维护策略[J]. 中国医疗设备, 2020, 35(11): 168-171.
LI Wenlian,ZHU Hui. Common Fault Analysis and Maintenance Strategy of Electronic Endoscope. China Medical Devices, 2020, 35(11): 168-171.