Abstract:Objective The existing automated instruments all use fixed hematocrit (HCT) value to convert the serum C-reactive protein (CRP) value. The serum CRP value needs manual reckoning if the hematocrit is not in normal. In this paper, we designed a kind of device which can test the hematocrit with automatic correction function. Methods The specific protein concentration and hematocrit were measured by scattering light: the linear relationship between CRP concentration and hematocrit, and the linear relationship between scattering light intensity and hematocrit were verified| specific protein concentration and hematocrit were measured by scattering method| correct the CRP results using whole blood hematocrit. Results There was a linear relationship between the whole blood CRP concentration and hematocrit, R2=0.9819, as well as between the scattering light intensity and hematocrit, R2=0.9672. This paper takes the sample whose hematocrit is 38% as the true value (49.15mg/L). All the relative deviations are less than 15% which was within the standard range. Conclusion The analyzer uses hematocrit to adjust the CRP value. The results show that this procedure is effectively.
陈月玲1,王永强2,郑继波3,陈兰兰4. 自动红细胞比容校正全血CRP测试结果特定蛋白分析仪的系统设计[J]. 中国医疗设备, 2020, 35(11): 80-84.
CHEN Yueling1,WANG Yongqiang2,ZHENG Jibo3,CHEN Lanlan4. Design for the Automated Protein Analyzer Using Hematocrit to Calibrate the Results of Whole Blood CRP. China Medical Devices, 2020, 35(11): 80-84.