Performance Analysis of A Thermoluminescent Dosimeter
ZHAO Jinpei1, WANG Linjie2, ZHOU Chao2, ZHANG Jian3, JIANG Qisheng4, LI Xiuqin1
1. Department of Health Protection, Beijing Military Center for Disease Control and Prevention, Beijing 100042, China;
2. Military Preventive Medical College, Army Medical University, Chongqing 400038, China;
3. Beijing JunYiLianRui Institute of Nuclear Measurement Technology, Beijing 100850, China;
4. Center Lab, General Hospital of Rocket Troops, Beijing 100088, China
Abstract:Objective To evaluate the reliability of a thermoluminescence dosimetry system by testing the key performance indexes.
Methods Under the criteria of JJG593-2006, GBZ207-2016, and GB/T10264-2014, eight indexes of the thermoluminescence dosimetry
system were examined, including the batch homogeneity, stability, angle response, linearity, energy response, quantity inspection,
residual dose, and the minimal detection limit. Moreover, the results were compared with the related standards. Results The results of
various performance indexes met the criteria of the national and industrial standards. Conclusion The thermoluminescence dosimetry
system meets the technical requirements of relevant standards, and can be used for personal and environmental dose monitoring.
赵进沛1,王琳杰2,周超2,张建3,江其生4,李秀芹1. 一种热释光剂量计的性能检测分析[J]. 中国医疗设备, 2019, 34(3): 37-40.
ZHAO Jinpei1, WANG Linjie2, ZHOU Chao2, ZHANG Jian3, JIANG Qisheng4, LI Xiuqin1. Performance Analysis of A Thermoluminescent Dosimeter. China Medical Devices, 2019, 34(3): 37-40.