Abstract:It is well known that blood analyzer plays an important role in clinical examination. The failure of blood analyzer will not
only affect the doctor’s diagnosis, but also bring unnecessary repeated examination to the patient, resulting in the economic loss of
patients. Two typical faults of Sysmex automatic blood analyzer were summarized in this paper. The first one was the error of blood
count, and the second one was positive negative pressure input anomaly. Through the combination of theoretical knowledge, we
conducted a step-by-step investigation starting from the detection of the fault alarm, and the failure was settled finally. This article
can provide references for peers. In the case of similar faults, it can be used to solve the problem quickly, shorten the time of failure,
and reduce the loss of hospital and patients.
丛宁宁. 希森美康全自动血液分析仪故障维修2例[J]. 中国医疗设备, 2018, 33(6): 95-97.
CONG Ningning. Two Cases of Failure Maintenance of Sysmex Automatic Blood Analyzer. China Medical Devices, 2018, 33(6): 95-97.