Abstract:According to the troubleshooting course of a common fault case of Siemens Primus Linear Accelerator Multi-leaf
Collimator (MLC) interlock fault, this paper can offer reference for the maintenance of linear accelerator. The leaves were inspected
one by one to find the cause of the malfunction, then, the drive gear positions of the blades were interchanged and one wear pads
with thickness of 1.5 mm were placed between the blade gear to reduce moving clearance caused by bearing damage. MLC runs
normally within one year inspection, this method is feasible for such malfunction, and can reduce the maintenance cost.
陈利军,谷晓华,杨留勤. Siemens Primus直线加速器多叶光栅故障排除一例[J]. 中国医疗设备, 2017, 32(9): 93-95.
CHEN Lijun, GU Xiaohua, YANG Liuqin. One Case of Troubleshooting for Siemens Primus Linear Accelerator Multi-Leaf Collimator. China Medical Devices, 2017, 32(9): 93-95.