Abstract:Regular electrical safety testing is an essential part in quality control of all in-use medical
equipment. In this paper, 50 infant incubators were tested by using ESA620 automatic electrical safety
analyzer based on the standard of IEC60601-1, including grounding resistance protection, insulation
resistance, earth leakage current, shell leakage current and other testing items. It could be seen from the
testing results that only the grounding resistance protection met the standard (the up-to-standard rate
was 70.0%). The main potential safety problems of the infant incubators were the deteriorated condition
of the power cords and the oxidization of the outermost shell. According to the post-maintenance test
results, the pass rate was 100%.
吴蕴蕴,沈云明,苏畅,王琦,
郑苔施. 婴儿培养箱的电气安全检测[J]. 中国医疗设备, 2016, 31(9): 124-127.
WU Yun-yun, SHEN Yun-ming,
SU Chang, WANG Qi,
ZHENG Tai-shi. Electrical Safety Testing for the Infant Incubator. China Medical Devices, 2016, 31(9): 124-127.