Structural Principle and Fault Analysis of Italian Opera FP Digital Photography System
ZHONG Liming1, HUANG Peirong2, DENG Zijie1, CHE Zehong1, LI Tao1
1. Department of Medical Equipment, Guidong People’s Hospital of Guangxi Zhuang Autonomous Region, Wuzhou Guangxi
543001, China; 2. Department of Equipment, People’s Hospital of Teng County, Guangxi Zhuang Autonomous Region, Wuzhou
Guangxi 543300, China
Abstract:Opera FP is a powerful and versatile all-digital photography system, which has a large number of users in China. This
paper expounded the basic structure of the system principle and main components. The fault code “099” in everyday use, the main
system initialization cannot complete, automatic shutdown and restart irregular high-voltage X-ray generator were troubleshot for
detailed analysis. And the corresponding repair measures were put forward for troubleshooting and method, by minimizing the
components level maintenance, replacement of capacitance, hard drives, optical fiber and regulator tube respectively, the faults were
ruled out one by one. The maintenance and troubleshooting for the majority of medical professionals counterparts have provided the
certain theory reference and practice experience.
钟里明1,黄培荣2,邓茈杰1,车泽宏1,李涛1. 意大利OPERA FP数字摄影系统结构原理及故障分析[J]. 中国医疗设备, 2022, 37(4): 177-180.
ZHONG Liming1, HUANG Peirong2, DENG Zijie1, CHE Zehong1, LI Tao1. Structural Principle and Fault Analysis of Italian Opera FP Digital Photography System. China Medical Devices, 2022, 37(4): 177-180.