Research on Intelligent Quality Control Experiment of CT Image Performance
ZHONG Yutong1, LI Mei1, LIU Shizhu2, ZHANG Jian3, JIAO Long2, YU Xikun2, LIU Jingxin4
1. College of Electronic Information Engineering, Changchun University of Science and Technology, Changchun Jilin 130028,
China; 2. Jilin Medical Imaging Engineering Technology Research Center, Changchun Jilin 130033, China; 3. Changchun City
Metrology and Testing Technology Research Institute, Changchun Jilin 130012, China; 4. Department of Radiology, China-Japan
Union Hospital of Jilin University, Changchun Jilin 130033, China
Abstract:Objective To study the method and feasibility of the computer judgment and analysis method for the image of CT quality
control phantom performance test, and compare it with the conventional method. Methods Scanning CT performance phantom,
the spatial resolution, low contrast resolution, noise, CT value and uniformity were measured by computer automatic analysis and
conventional detection method. Results The results showed that the performance index of CT was judged based on computer
software and conventional testing method. The results were consistent between software measurement and traditional measurement,
and the method of CT intelligent quality control based on computer software was feasible. Conclusion Through the intelligent
measurement of CT image performance index, it can evaluate the performance of CT equipment more objectively and accurately,
reduce manual error, and improves the efficiency of CT quality control greatly.
钟宇彤1,李梅1,刘师竺2,张健3,焦龙2,于喜坤2,刘景鑫4. CT影像性能智能质控实验研究[J]. 中国医疗设备, 2019, 34(4): 1-4.
ZHONG Yutong1, LI Mei1, LIU Shizhu2, ZHANG Jian3, JIAO Long2, YU Xikun2, LIU Jingxin4. Research on Intelligent Quality Control Experiment of CT Image Performance. China Medical Devices, 2019, 34(4): 1-4.