Experimental Research for DR Imaging Quality Control test
LIU Yuchen1,2, YU Xikun2, LIU Shizhu2, JIAO Long2, ZHANG Jian3, LIU Jingxin1,2
1. Department of Radiology, China-Japan Union Hospital of Jilin University, Changchun Jilin 130033, China; 2. Jilin Medical
Imaging Engineering Technology Research Center, Changchun Jilin 130033, China; 3. Changchun City Metrology and Testing
Technology Research Institute, Changchun Jilin 130033, China
Abstract:Objective To study which DR performance parameters should be frequently tested by medical institutions, so as to realize
mutual recognition and sharing of medical images Methods Some common DR equipment that was normally used in medical
institutions were selected, and relevant performance parameters were tested referring to the national metrological verification
regulation JJG 1078-2012 medical digital photography (CR, DR) system X-ray radiation source for testing relevant performance
parameters. Then, statistical analysis was conducted on the data. Results The main performance parameters that affected the quality
of DR imaging were spatial resolution and low contrast resolution. Conclusion In order to realize mutual recognition and sharing of
medical images, it is suggested that medical institutions should often test the spatial resolution and the low-contrast resolution.
刘宇辰1,2,刘师竺2,于喜坤2,焦龙2,张健3,刘景鑫1,2. DR成像质量及影响因素的检测分析[J]. 中国医疗设备, 2019, 34(4): 18-20.
LIU Yuchen1,2, YU Xikun2, LIU Shizhu2, JIAO Long2, ZHANG Jian3, LIU Jingxin1,2. Experimental Research for DR Imaging Quality Control test. China Medical Devices, 2019, 34(4): 18-20.