Abstract:Objective This study aimed to compare the performance of dynamic ECG recording box
(Baihui CT-08) and conventional ECG machine in premature beat acquisition. Methods Premature beat
signal from a physiological signal simulator was simultaneously collected via the 2 devices. A total of 72
sets of signal were obtained and statistically analyzed to verify the consistency of the 2 devices. Results
No statistically significant difference was found between the 2 devices in the function of total beats and
premature beats recording on all types of heart disorders except for the type of “PVC1 R on T, ventriculus
sinister”. Conclusion CT-08 is highly consistent with Holter in premature beat recording, making it
possible to be promoted as a demonstration product in the project of national innovative medical devices
and technical achievements transformation.
毛彬,孙静,郑骏,陈华,
王吉鸣,冯靖祎. 国产动态心电记录仪CT-08的临床应用评价[J]. 中国医疗设备, 2017, 32(5): 98-101.
MAO Bin, SUN Jing,
ZHENG Jun, CHEN Hua,
WANG Ji-ming, FENG Jing-yi. Evaluation on Clinical Application of Dynamic ECG Recording Box CT-08. China Medical Devices, 2017, 32(5): 98-101.