Abstract:In reference to Calibration Specification for Defibrillator Monitor, this paper adopted the Fluke
Impluse7000DP Defibrillator Analyzers to conduct quality control to test five parameters of cardiac defibrillator
monitors in our hospital, including appearance and function, delivered energy, charging time,
delay time under synchronizer, and heart rate value. Timely maintenance was carried out on defibrillators
with certain problems to ensure the safe use of defibrillators in clinical practice. Meanwhile a comparative
analysis was conducted on the results of quality control. The research explored a quality control
program that was appropriate for defibrillator quality control in the context of clinical experience in the
hospital.
朱悦,尹钢,廖云粤. 除颤监护器质量控制探讨[J]. 中国医疗设备, 2016, 31(5): 113-116.
ZHU Yue, YIN Gang,
LIAO Yun-yue. Discussion on Quality Control of Defibrillator Monitor. China Medical Devices, 2016, 31(5): 113-116.