Development and Verification of Sputum Suction Device Based on Magnetic Coupling Drive
TAN Wenjun1a,1b, WANG Saiyu2, XING Binyu1a, TANG Bo1b, YUAN Lirong1b, LV Yi1b,1c, ZHU Haoyang1b,1d
1. a. Department of Surgery Intensive Care Unit; b. National Local Joint Engineering Research Center for Precision Surgery &
Regenerative Medicine; c. Department of Hepatobiliary Surgery; d. Department of Anesthesiology, The First Affiliated Hospital
of Xi’an Jiaotong University, Xi’an Shaanxi 710061, China; 2. Department of Nursing, Xi’an Jiaotong University Health Science
Center, Xi’an Shaanxi 710048, China
Abstract:Objective To develop a new generation of stable and safe sputum suction device to replace nursing staff in sputum suction
operation, aiming at the core operation of the delivery tube and the withdrawal tube in the artificial airway of the previous sputum
suction device. Methods Based on the research and development of the previous generation of simulated sputum suction robot,
the push-pull, speed, time and other relevant parameters of the sputum delivery and desorption tube had been obtained, by using
magnetic coupling drive and roller transportation, non-contact power transmission was achieved to reduce the holdup and ensure the
stability of the sputum suction tube delivery and withdrawal. Through the feasibility training on the sputum suction practice model
of artificial airway, the effectiveness and safety of the sputum suction device were ensured. Results Feasibility training showed that
in 20 effectiveness simulation experiments, about 5-9 mL of simulated sputum was successfully extracted in 17 experiments, with an
average extraction of about 7.5 mL. In 10 safety simulation experiments, there was no significant difference in the thrust, tension and
rotation force between sputum suction device and manual sputum suction device in the process of delivery and withdrawal (P>0.05).
The data of sputum suction device was more stable. Conclusion The sputum suction device can transmit and exit the sputum suction
tube stably and safely, which can effectively complete the sputum suction operation. It is easy to disassemble, clean and disinfect,
which can effectively reduce the work burden of medical staff.
谭文君1a,1b,王赛宇2,邢斌瑜1a,汤博1b,袁力蓉1b,吕毅1b,1c,朱皓阳1b,1d. 基于磁耦合传动的吸痰装置的研制与验证[J]. 中国医疗设备, 2024, 39(3): 20-25.
TAN Wenjun1a,1b, WANG Saiyu2, XING Binyu1a, TANG Bo1b, YUAN Lirong1b, LV Yi1b,1c, ZHU Haoyang1b,1d. Development and Verification of Sputum Suction Device Based on Magnetic Coupling Drive. China Medical Devices, 2024, 39(3): 20-25.