Abstract:By analyzing the frequent failure of Baxter jinbao AK96 haemodialysis machine in function self-test, the experience was
provided for haemodialysis machine maintenance in this study. For C COFF 088 058 alarm in function self-check, the root cause of
its failure was analyzed, the default disinfection program was modified to ensure the safety and effectiveness of dialysis machine
disinfection, the position of the machine was adjusted, and the maintenance interval of air filter was shortened. After half a year of
operation monitoring, although the failure still occurs very rarely, it is not enough to cause trouble to clinical work. In view of this
kind of fault phenomenon, this maintenance method is feasible
张旭林. 一种百特金宝AK96血液透析机频发故障分析与排除[J]. 中国医疗设备, 2019, 34(8): 183-185.
ZHANG Xulin. Analysis and Elimination of A Frequent Fault of Baxter Jinbao AK96 Hemodialysis Machine. China Medical Devices, 2019, 34(8): 183-185.