Abstract:Objective To reduce the occurence rate of failures, the maintenance experience was summarized through ananlyzing and
repairing the “no count” fault of SIENENS ECAMHD3R ECT probe. Methods Probe 1 and probe 2 on the collimator were measured
by point source test. Then fault points were found through multiple sets of tests after entering the local service mode of the machine.
Results After maintenance, it was found that the failure component was PI59R plate of probe 2, and the machine returned to normal
use after replacement. Conclusion The structure ECAMHD3R ECT probe was complicated, there were lots of difficult troubles.
Maintenance of ECAMHD3R ECT probe should be carried out based on detailed analyzing the operating principle of the equipment
and finding faults carefully.
孙传金,吉蘅山,王新刚,朱虹. 西门子ECAMHD3R探头“无计数”的维修实例及经验总结[J]. 中国医疗设备, 2019, 34(1): 169-170.
SUN Chuanjin, JI Hengshan, WANG Xingang, ZHU Hong. Maintenance Examples and Experience Summary of Siemens ECAMHD3R Detector “No Count”. China Medical Devices, 2019, 34(1): 169-170.