Abstract:In this paper, the working principle of philips color ultrasound named by EPIQ7C and the software and hardware process
of power startup were introduced in detail firstly. Then, the detection of power failure was mainly discussed in combination with the
equipment failure condition. By analyzing the voltage threshold of power module interface connection diagram and power module,
the isolation detection method of power failure was discussed. It provided strong data and theoretical support for the power modular
maintenance of EPIQ7C.
李玲,时先锋,董冲,李凯凡,谭绍东,苏晓舟. 飞利浦彩超EPIQ7C结构原理及电源故障检测[J]. 中国医疗设备, 2018, 33(7): 91-93.
LI Ling, SHI Xianfeng, DONG Chong, LI Kaifan, TAN Shaodong, SU Xiaozhou. Principle and Power Failure Detection of Philips Color Ultrasound EPIQ7C. China Medical Devices, 2018, 33(7): 91-93.