Abstract:Objective Taking the testing of the synthetical relative effect of edge light energy as an example, this paper discussed the
uncertainty analysis of endoscope testing device in combination with the feature of medical rigid optical endoscope testing. Methods
According to the uncertainty determination of A and B type, the main affecting factors were analyzed and the key parameters of the
testing method were extracted. We compared the uncertainty measurement results of ordinary endoscope testing device and integrated
endoscope testing system. Results The essential designing elements of endoscope testing devices were provided. Conclusion Our
results provide research basis for further improvement of endoscope testing accuracy.
孟祥峰,王浩,刘艳珍,任海萍. 使用集成装置进行医用硬性光学内窥镜检测的不确定度分析[J]. 中国医疗设备, 2017, 32(8): 16-19.