Abstract:Objective To explore the feasibility of EIT-based (Electrical Impedance Tomography) real-time monitoring of the little pig pulmonary hemorrhage model. Methods As the objective of this study, electrical impedance changes of little pig pulmonary hemorrhage model was constantly and real-timely monitored with adoption of EIT. Results Gradual increase of each 5 mL of injected blood volume resulted in redness displayed in EIT images incorresponding areas and gradual enhancement. The hemorrhage locations revealed by EIT accorded with those displayed by CT. Quantitative analysis results indicated that gradual increase of each 5 mL of injected blood volume leaded togradually-enlarge delectrical impedance and area in the relevant locations in EIT images, which proved the relevance between EIT quantitative analysis indicators and the injected blood volume. Conclusion EIT could real-timely monitor the electrical impedance changes caused by pulmonary hemorrhage in little pigs in an effective way, which laid a solid foundation for clinical researches.
代萌,徐灿华,杨滨,夏军营,刘本源,刘锐岗,史学涛,尤富生,付峰. 基于电阻抗成像技术实时监测小猪肺出血模型的实验研究[J]. 中国医疗设备, 2015, 30(7): 19-21.
DAI Meng, XU Can-hua, YANG Bin, XIA Jun-ying, LIU Ben-yuan, LIU Rui-gang, SHI Xue-tao, YOU Fu-sheng, FU Feng. Experiment Study on EIT-Based Real-Time Monitoring of the Little Pig Pulmonary Hemorrhage Model. China Medical Devices, 2015, 30(7): 19-21.