Fault Analysis and Maintenance Improvement of ECG-1350C Multichannel Electrocardiograph
QIAO Min1, LIU Weijun2, WANG Yajun3, XU Wei1, DONG Haokun1, MI Xingyu4
1. Department of Medical Equipment, Baoji Central Hospital, Baoji Shaanxi 721008, China;
2. Department of Medical Equipment, Shaanxi Provincial People’s Hospital, Xi’an Shaanxi 710068, China;
3. Department of Medical Equipment, Shaanxi Provincial Tumor Hospital, Xi’an Shaanxi 710061, China;
4. Department of Equipment, Tongchuan Taoyuan Community Staff Hospital, Tongchuan Shaanxi 727000, China
Abstract:To provide reference for the troubleshooting of ECG-1350C multichannel electrocardiograph, the equipment principle and
clinical application functions were described in detail, and three typical practical maintenance cases were used for troubleshooting. The
paper-feeding motor of the multi-channel electrocardiograph was improved according to the observation method, replacement method,
meanwhile, the test method and SD wireless LAN card, and the USB interface slot were repaired and improved. By fully understanding
of the principle of multichannel electrocardiograph and the use of various clinical application functions, the fault caused by the paper
motor, SD wireless LAN card, and USB interface slot was repaired. The repair of the three faults was summarized and analyzed. Based
on this, improving the design defects of the device can save maintenance costs and reduce the subsequent failure rate of the device.
乔敏1,刘伟军2,王雅军3,徐伟1,董昊坤1,米星宇4. ECG-1350C多道心电图机常见故障分析及维修改进探讨[J]. 中国医疗设备, 2019, 34(12): 169-172.
QIAO Min1, LIU Weijun2, WANG Yajun3, XU Wei1, DONG Haokun1, MI Xingyu4. Fault Analysis and Maintenance Improvement of ECG-1350C Multichannel Electrocardiograph. China Medical Devices, 2019, 34(12): 169-172.