Three Maintenance Cases of GE Voluson E8 Color Ultrasonic Diagnostic Instrument
YU Mingtao1, YU Liang2, CHEN Jie3
1. Department of Equipment, Yantaishan Hospital, Yantai Shandong 264008, China; 2. Department of Equipment, Fushan District
People’s Hospital of Yantai City, Yantai Shandong 265599, China; 3. Sterilization and Supply Center, Yantai Yuhuangding Hospital,
Yantai Shandong 264099, China
Abstract:GE Voluson E8 is an advanced real-time three-dimensional color ultrasound diagnostic instrument, which plays an
important role in medical activities. In this paper, the imaging principle of GE Voluson E8 was summarized, the fault location was
explored through different troubleshooting methods, and 3 cases of image display failure and startup failure were introduced. The
fault phenomena were the problem of image display, the failure to turn on the display after screen crash and forced shutdown, and the
pencil-shaped high-echo interference in the image. The fault spots were determined as graphics card, graphics card and RTN main
power module, RSR and RST transmitting and receiving board by step investigation, and the fault was finally eliminated by replacing
corresponding components. The three maintenance cases introduced in this paper suggest that the troubleshooting should bedone
from easy to difficult , so as to save the maintenance time and find the fault point more accurately.
于明涛1,于良2,陈杰3. GE Voluson E8 彩色超声诊断仪故障维修3例[J]. 中国医疗设备, 2022, 37(7): 178-169.
YU Mingtao1, YU Liang2, CHEN Jie3. Three Maintenance Cases of GE Voluson E8 Color Ultrasonic Diagnostic Instrument. China Medical Devices, 2022, 37(7): 178-169.