Abstract:This paper mainly introduces the tissue cutting principle of ERBEJET2 waterjet applied in clinical practice as well as the troubleshooting methods and maintenance of common faults. According to the analysis of ERBE after-sales maintenance cases in
recent three years (2016—2018), the troubleshooting and maintenance ideas of common faults were refined. Some hardware faults and common error codes of ERBEJET2 waterjet were analyzed. And the faults were repaired by cleaning the pump set module,
replacing the main chip, and replacing the diode, and summarizing the maintenance experience finally. Understanding the principle and basic structure of ERBEJET2 waterjet can effectively improve the utilization rate of equipment, reduce the occurrence of
common faults and promote the quality of after-sales maintenance.
丁辉,陆晓华,宋海峰. ERBEJET2水刀的原理及常见故障的分析维修[J]. 中国医疗设备, 2020, 35(12): 184-187.