Abstract:The quality control of the dental panoramic X-ray machine is of great significance as it is an important diagnostic device in stomatology medical domain. Some explorations to measure quality control of the dental panoramic X-ray machine are carried out. This paper mainly introduces quality control methods of some technical parameters, such as radiation dose, the half-value layer, spatial resolution, low-contrast resolution, tube voltage and exposure time. In order to adapt to imaging characteristics of the dental panoramic X-ray machine, the ‘U’ type low-contrast resolution phantom has been designed.
姚绍卫,夏勋荣,徐睿智. 口腔全景X射线机的质量控制[J]. 中国医疗设备, 2013, 28(7): 47-49.
YAO Shao-wei, XIA Xun-rong, XU Rui-zhi. Quality Control for Dental Panoramic X-ray Unit Machine. China Medical Devices, 2013, 28(7): 47-49.