Abstract:Quality control of multi-parameter monitor is conducted to ensure the efficacy and safety of the machine. Analysis of the testing data shows that oxidation of the external terminal and aging of the power cord are the main causes of big resistance value of ground impedance. The problems of non-invasive blood pressure and respiratory rate indicate that quality control should be verified through a variety of textual researches. Testing equipments which will be used in clinical quality control should be tested by quality department.
王守镜,徐林,王梦婷,杨罗宽. 多参数监护仪的质量控制分析[J]. 中国医疗设备, 2013, 28(10): 62-63.
WANG Shou-jing, XU Lin, WANG Meng-ting, YANG Luo-kuan. Analysis of Quality Control of Multi-parameter Monitor. China Medical Devices, 2013, 28(10): 62-63.