Abstract:According to the transformation relationship between the LAP DORADO 3 laser system coordinates and treatment coordinates, a specialized Excel worksheet was designed by using the formula editing function. The final position data of the LAP laser lights and CT scanning table can be obtained automatically, as long as corresponding raw data are input into the specialized Excel worksheet. Then the laser lights and CT scanning table can be driven to shift with the final position data, and the treatment isocenter for patients with radiotherapy can be determined. This paper introduces the application method for determining treatment isocenter for patients with radiotherapy, by using the specialized Excel worksheet. It can reduce the manual calculation workload, avoid occurrence of manual calculation errors, and improve the work efficiency effectively, by using the above worksheet.
时飞跃,韩晶晶,秦伟,陈成,陈丽丽. 用Excel工作表确定患者治疗等中心的应用方法[J]. 中国医疗设备, 2013, 28(12): 12-14.