Abstract:With the enforcement of YY 0505-2012, the monitoring method of electromagnetic compatibility (EMS) testing for medical electrical equipment should be studied. Based on the requirements of relevant standards, this paper discusses the essential performance of medical electrical equipment. Taking microwave therapy equipment as an example, it discusses the confirmation and monitoring methods of the essential performance in EMC immunity test.
王占硕,王颖,高山,杨建刚. 医用电气设备电磁兼容检测中基本性能的确定及其监测方法初探[J]. 中国医疗设备, 2014, 29(2): 20-23.
WANG Zhan-shuo, WANG Ying, GAO Shan, YANG Jian-gang. Confirmation and Monitoring of Essential Performance of Medical Electrical Equipment in Electromagnetic Compatibility Test. China Medical Devices, 2014, 29(2): 20-23.